9408-80 Flow-Through Dissoved Oxygen System
Brand: ABB GroupThe Dissolved Oxygen System Cartridge 9408-80 is a compact and robust flow-through system used to measure high levels of dissolved oxygen in process and waste waters.
The sensor is extremely rugged and can be used in the most demanding environments.
9408-80 systems complement the highly successful 9408 floating ball, dip and submersible systems and employ the well-proven 8012 oxygen sensor.
A key feature is the bayonet snap-fit connection which makes removal for calibration very easy and convenient.
The typical life of the 8012 oxygen sensor is 12 months, although many users benefit from a much longer operational period.
Features- Used up to 20 ppm or 200 % saturation
- Bayonet-fit connection makes removal simple
- Choice of process connections in BSP or NPT
- Replaceable oxygen sensor simplifies maintenance
- Water-wash option - keeps the sensor in optimum condition
Specifications
Brand
Accuracy
- ± 0.2 mg/L or ± 2 % saturation within ± 10° C of the calibration Temperature within the Range 0° to 35° C
Range
- Dissolved oxygen: 0 to 20 ppm
Material of Construction
- Glass-Coupled Polypropylene
Operating Temperature
- -5° to 40° C
Operating Pressure
- Atmospheric
Response Time
- Typically 20 seconds for 90 % step change of oxygen concentration at 20° C
Pressure Rating
- 2 bar
Flow Rate
- 2 L/min
Process Connection
- 1/2" BSP
- 1" BSP
- 1/2" NPT
- 1" NPT
Cable Length
- Standard 1 to 20 m
- Maximum 100 m with junction box